FLUXUS WW A Menu structure
132 UMFLUXUS_WWV1-1EN, 2018-02-28
Special Funct.\SYSTEM settings\Measuring\Calibrat. data
selection of the measuring channel whose flow parameters are to be de-
fined
This display will only be indicated if the SuperUser mode is activated.
determination of the profile bounds
factory: the default profile bounds will be used
user: the profile bounds can be defined
This display will only be indicated if the SuperUser mode is activated.
input of the max Reynolds number at which the flow is laminar
range: 0...25 500 (rounded to hundreds)
0: the default limit of 1 000 will be used
This display will only be indicated if the SuperUser mode is activated and
Profile bounds = user is selected.
input of the min. Reynolds number at which the flow is turbulent
range: 0...25 500 (rounded to hundreds)
0: the default limit of 3 000 will be used
This display will only be indicated if the SuperUser mode is activated and
Profile bounds = user is selected.
request if an additional correction of the flow velocity is to be defined
on: the correction data can be defined
off: without correction of the flow velocity
This display will only be indicated if the SuperUser mode is activated.
input of the slope for the correction formula
range: -2.000...+2.000
0: without correction
This display will only be indicated if the SuperUser mode is activated and
Calibration = on is selected.
input of the offset
range: -12.7…+12.7 cm/s
0: without offset
This display will only be indicated if the SuperUser mode is activated and
Calibration = on is selected.
SYSTEM settings\Proc. outputs
selection of the displays for the transmitter outputs setting
selection of the output to be installed
SYSTEM settings\Storing
selection of the displays for the storing of measured values in the data
logger
setting of the data logger overflow behavior x
INIT-resis-
tant
Calibrat. data
for Channel A:
A:Profile bounds
factory >USER<
Laminar flow
ifR*< 0
Turbulent flow
if R*> 0
A:Calibration ?
>OFF< on
A:Slope=
1.000
A:Offset=
0.0 cm/s
SYSTEM settings
Proc. outputs
Install Output
Current I1
SYSTEM settings
Storing
Ringbuffer
off >ON<