A Menu structure FLUXUS WW
UMFLUXUS_WWV1-1EN, 2018-02-28 133
setting of the storing behavior of the totalizers
• one: the value of the totalizer that is currently displayed will be stored
• both: one value for each flow direction will be stored
x
activation of the signal amplitude storing
The value will only be stored if the data logger is activated.
x
activation of the fluid sound speed storing
The value will only be stored if the data logger is activated.
x
activation of storing of the diagnostic values x
activation of an acoustic signal every time a measured value is stored or
transmitted
x
SYSTEM settings\serial transmis.
selection of the displays for the format of the serial transmission of mea-
sured values
activation of the serial transmission with/without blanks x
selection of the decimal marker for floating point numbers x
selection of the character for column separation x
selection of the serial interface
default: RS232
This display will only be indicated if the transmitter has an RS485 inter-
face.
x
SYSTEM settings\Network
change of settings for the transmission parameters of the RS485 interface
input of the address of the measuring instrument x
confirmation or change of the transmission parameters x
change of the baud rate, parity or number of stop bits x
INIT-resis-
tant
Quantity Storage
one >BOTH<
Store diagnostic
off >ON<
SYSTEM settings↕
serial transmis.
SER:decimalpoint
’.’ >’,’<
SER:col-separat.
’;’ >’TAB’<
Send Offline via
RS232 >RS485<
RS485 protocol
default >SETUP<
>BAUD< parity st
1200 EVEN 1