13 Outputs
13.1 Analog output PIOX R721
2021-09-01, UMPIOX_F72xV1-3EN
70
13.1.4 Output of a status value/event value
• Select the list item Status in the menu item Source
item.
• Press ENTER.
Output range
• Select a list item.
– 4...20 mA
– Other range
• Press ENTER.
If Other range is selected, enter the values Output
MIN and Output MAX.
The output range has to be > 10 % of the max. output value (Output MAX). An error message will be displayed if the
output range is smaller. The next possible value will be displayed.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing or further information is displayed.
• Press ENTER.
Output function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select Yes to test the output. Select No to display the next menu item.
• Press ENTER.
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select Repeat to repeat the test, Finish to display the next menu item.
• Press ENTER.
Source item
Status
Output range
Status OK Idle state
Output info
Test signal
Test mea. range
see annex A
Outputs\...\Output range
status value – state OK event value – idle state
The status of the output signal is defined which is to be output
when measuring a measured value.
The status of the output signal is defined which is to be output if
no event occurs.
• Select in the scroll list the value for state OK.
• Press ENTER.
• Select the value for the idle state.
• Press ENTER.
Outputs\...\Output info
Outputs\...\Test signal
Outputs\...\Enter test value