EasyManua.ls Logo

Fluke 1672 - Page 48

Fluke 1672
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1672/1673 FC/1674 FC
Users Manual
42
4. Select RCD current setting: 10 mA, 30 mA, 100 mA, 300 mA, 500 mA or Var.
5. Tap
H.
6. Select the RCD current setting (10, 30, 100, 300, 500, or 1000 mA).
7. Select a test current multiplier (x ½, x 1, x 5, or Auto). Normally you will use x 1 for this test.
8. Select the RCD type. See Figure 16.
Figure 16. RCD Type Selection
AC current to test type AC (standard AC RCD) and type A (pulse-DC sensitive RCD)
Half-wave current to test type A (pulse-DC sensitive RCD)
Delayed response to test S-type AC (time delayed AC RCD)
Delayed response to S-type A (time delayed pulse-DC sensitive RCD)
1673 FC/1674 FC
Smooth-DC current to test type B RCD
Delayed response to S-type B (time delayed smooth-DC current RCD)
Note
For type G, K or R RCDs, choose type A (half-wave current). Pass is not triggered on the
short delay of 10 ms of the G, K, and R types. These types need a trip time of at least
10 ms.
Type B+ RCDs are tested with type B smooth dc-current.
Form
Functions
No limit
No project selected

Table of Contents