Service Manual
1-14
Environmental
Warmup Time
1 hour to rated specifications
Temperature Coefficient
<0.1 times the applicable accuracy specification per degree C for 0°C to 18°C and 28°C
to 50°C
Electromagnetic Compatibility
In an RF field of 1 V/m on all ranges and functions:
Total Accuracy = Specified Accuracy + .4% of range. Performance above 1 V/m is
not specified.
Operating Temperature
0°C to 50°C
Storage Temperature
40°C to + 70°C
Elevated temperature storage of battery will accelerate battery self-discharge. Maximum
storage time before battery must be recharged:
20 - 25°C 1000 days
50°C 180 days
70°C 40 days
Relative Humidity
To 90% at 0°C to 28°C , (non condensing)
To 80% at 28°C to 35°C,
To 70% at 35°C to 50°C except to 70%
at 0°C to 50°C for the 1000 k, 3 M, 10 M, 30 M, 100 M, and 300 M
ranges.
Altitude
Operating 0 to 10,000 feet
Non-operating 0 to 40,000 feet
Vibration
3 G @ 55 Hz
Shock
Half sine 40G. Per Mil-T- 28800D, Class 3, Style E.
Bench Handling. Per Mil-T-28800D, Class 3.