Introduction and Specifications
Specifications
1
1-15
General
Common Mode Voltage
1000 V dc or rms ac maximum from any input to earth
Size
9.3 cm high, 21.6 cm wide, 28.6 cm deep
Weight
Net, 2.4 kg without battery; 3.2 kg with battery; Shipping, 4.0 kg without battery, 4.8
with battery.
Power
90 to 264 V ac (no switching required), 50 and 60 Hz. 15 VA maximum
Safety
Compliant with the following standards:
ANSI/ISA S82.01-1994
CAN/CSA-C22.2 No. 1010. 1-92
EN61010.1:1993 to 1000 V Overvoltage Cat I, 600 V Overvoltage Cat II
UL3111-1. See the following for an explanation of CATI and CATII categories.
CAT I:
OVERVOLTAGE (Installation) CATAGORY I, Pollution Degree 2 per IEC1010-1
refers to the level of Impulse Withstand Voltage protection provided. Equipment of
OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which
measures are taken to limit the transient over voltages to an appropriate low level.
Examples include protect electronic circuits.
CAT II:
OVERVOLTAGE (Installation) CATAGORY II, Pollution Degree 2 per IEC1010-1
refers to the level of Impulse Withstand Voltage protection provided. Equipment of
OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied
from the fixed installation. Examples include household, office, and laboratory
appliances.
Electromagnetic Compatibility
Meets FCC Part 15 Subpart J.
EN61326-1 (1998)
RS-232-C
Baud rates:
300, 600, 1200, 2400, 4800 and 9600
Odd, even or no parity One stop bit