45
Service Manual
1-16
Options
Battery (Option -01K)
Type
8 V, Lead-Acid
Operating Time
8 hours (typical). N lights when less than 1/2 hour of battery operation remains. Meter
still meets specifications.
Recharge Time
16 hours (typical) with meter turned off and plugged into line power. Battery will not
charge when meter is turned on.
IEEE-488 (Option -05K)
Capability Codes
SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, E1, TE0, LE0 and C0 External Trigger
Input VIH 1.35 V minimum VIL 1.25 V maximum
Input Threshold Hysteresis
0.6 V minimum