5700A/5720A Series II Calibrator
Service Manual
3-12
RESISTANCE STANDARD
HI
LO
HI
LO
SENSECURRENT
HI
HI
LO
LO
HI
OUTPUT
V A
SENSE
V
AUX
CURRENT
GUARD GROUND
WIDEBAND
CHASSIS
GROUND
CALIBRATOR
F7-4.EPS
Figure 3-4. 742A-1 and 742A-10k External Calibration Connections
10. To reject the changes, return to the calibration menu shown in step 2 by pressing
P. Otherwise, press the softkey under “Proceed” to accept and save the changes,
and to open the display shown below, letting you continue with calibration.
PREV
MENU
IF3-12.EPS