Section 4: Using the Model 9100: Logic-Pulses Function 4.16-5
Final Width = 215mm
4.16.4.2 'Pulse Width' and 'Repetition' Time-Interval Editing
Pulse Width and Repetition time-intervals can be changed using 'Digit' and 'Direct' edit
facilities. The editing processes are not described in Section 3, but follow the same
general rules as for editing voltages.
The resolution of time-interval values is set at six significant digits, except for the shortest
span of intervals at four significant digits. This leads to four interval spans. The '
Pulse
Width
' interval value must always be at least 0.3µs shorter than the 'Repetition' interval.
The thresholds between resolutions of Pulse Width interval and of Repetition interval are
given in paras 4.16.5.1.
Hardware configurations for time-interval change are given in paras 4.16.5.2.