2-14 Section 2: Installing the Model 9100
Final Width = 215mm
2.8.5 Phase Lock Connections (Rear Panel)
These two BNC sockets are located at the lower center
of the rear panel, providing a reference phase output
(PHASE LOCK OUT) or synchronizing phase input
(PHASE LOCK IN) so that any 9100 can be used either
as the source, or the recipient, of a phase reference.
IN
OUT
PHASE
LOCK
The 9100 can be used in both roles simultaneously, being phase-synchronized to an
external input signal while passing on the phase of its analog output to another 9100.
Internal controls are provided (via front-panel keys or via the IEEE-488 / SCPI interface)
to adjust the phase-shift of the 9100 analog output and 'Phase Lock Out' signal with
respect to the incoming synchronizing 'Phase Lock In' signal.
Signal levels and loading are discussed in Section 4, Paras 4.4.3.5 (ACV) and Paras
4.6.3.5 (ACI).
2.8.6 'TRIG OUT' and 'SIG OUT' Connections (Rear Panel)
These two BNC sockets are located at the upper center of the rear panel, providing signal
and trigger outputs for the Oscilloscope Calibration Options 250 and 600.
Further details are discussed in Volume 3 of this handbook.