Section 4: Using the Model 9100: AC Harmonics Function 4.21-13
Final Width = 215mm
Sequence of Operations
Refer to the table of UUT calibration points in the UUT Manufacturers Calibration Guide.
Follow the correct sequence of calibration stages as directed by the guide, and carry out
the following operations (1) to (5) at each stage.
1. 9100 a. Use the front panel controls to set the 9100 Main Output to the
UUT cal point frequency, voltage and current, entering High
Voltage State if the cal point has been assigned to that State. The
default High Voltage State boundaries are shown in Fig. 4.21.1.
b. Select the required Harmonic parameters.
2. UUT Select the correct range for the cal point.
3. 9100 a. Set Output ON.
b. Slew the appropriate 9100 output reading until the UUT reading
is equal to the calibration point value.
4. UUT Record the 9100 screen output value as detailed in the UUT
Manufacturer’s Calibration Guide.
5. 9100 Set Output OFF.