This menu allows you to alternate between menu display 1 and 2 every few seconds.
Selections are: On or O
Press OK (F4) to accept selection. Press EXIT (F4) repeatedly until “Normal Mode” is seen briey to
exit the programming mode.
Password
There are two user level passwords, only Level 1 is programmable and gives access to all the normal
settings. The second password is used to allow access to calibration settings.
Default Level 1 password is “1234”, and Level 2 password is “9111”. The Level 1 programmable
password can be disabled by setting it to “0”.
From the SET PARAMETERS menu press DSP (F3) to select the DISPLAY menu.
To Program the Password
F1 F2 F3 F4
Press PSW (F3) key to select password.
F1 F2 F3 F4
This screen displays the current Level 1 password.
Press CHG (F1) key to change the password and enter new value.
Press OK (F4) to accept new data and exit programming by pressing EXIT (F4) key repeatedly until
out of the programming mode.
NOTE! Password can be number or letter characters up to 4 digits.
Units Settings
This menu is used to set the units for mass ow, temperature, and pressure as well as the setting
of reference temperature and reference pressure. These values will be set at Fox Thermal using
information supplied by the customer. These values can be changed to match a new application. The
units setting is accessed from the SET PARAMETERS menu. To access the Unit Settings Menu: