Temperature Units
After pressing OK (F4) to accept the Flow unit the display will prompt for the temperature unit
setting:
F1 F2 F3 F4
Press NXT (F1) to change selection and OK (F4) to accept.
Selections for Temperature units are: Deg C or Deg F
Reference Temperature
After pressing OK (F4) to accept the temperature unit setting, the display will prompt for temperature
reference in selected unit.
F1 F2 F3 F4
Press CHG (F1) to change the reference and OK (F4) to accept.
Pressure Units
After pressing OK (F4) to accept the reference temperature, the display will prompt for the reference
pressure unit selection:
F1 F2 F3 F4
Press NXT (F1) to select next entry and OK (F4) to accept.
Selections are:
mmHG = Millimeters of mercury (absolute)
Psia = Pounds per square inch absolute
bara = Bar absolute
Reference Pressure
After the pressure unit selection is made, the display will show a menu to enter the reference pressure:
F1 F2 F3 F4
Press CHG (F1) to change it and OK (F4) to accept.
TMP UNT= Deg F
NXT OK
TmpRef = 60 °F
CHG OK
PRES UNT= Psia
NXT OK
PresRef= 14.7
CHG OK
Operation | 66
Model FT3
Operation