e.bloxx A6-2CF
CONFIGURATION
46
HB_EBLOXX-A62CF_E_V19.doc
Gantner Instruments Test & Measurement GmbH
Example: Factor is 10 [kN] / 0.002 [V/V] = 5,000 the Offset is 0 kN
• Zeroing
This menu offers the selection whether the zeroing can be done by host and/or by digital inputs. If the option host is
selected, the zeroing can be done in the register card
Measure
. The zero value will always be stored non-volatile.
Pushing the button
Zeroing
this pop up menu will appear:
Example: Zeroing can be done from the host and by a digital signal on input DI 1.
• Tarring
This menu is comparable with the Zeroing menu, additional it is selectable whether the tare value will be stored non-
volatile. To store the value volatile is recommended when the tarring function will be used periodically in short cycles.
Example:
Tarring can be done from the host and by a digital signal on input DI 2.