PanaFlow™ LC User’s Manual 45
Chapter 3. Programming
Figure 57: Setting up Analog Output
3.5.2c Calibrating Analog Output
To calibrate Analog Output, use steps as in section “Log-in and Primary Pages” to navigate to the
Input/Output settings page. You may use a multimeter or DCS/SCADA to calibrate the Analog Output.
Regardless of whether multimeter or DCS/SCADA is used, the steps below remain the same. For better
readability, the steps below only indicate multimeter and does not repeat multimeter or DCS/SCADA.
1. Turn ON the multimeter (if used), and set it to measure Current (mA) DC. Connect test lead from
positive side (Aout+) of the Main Analog Output to the positive terminal of multimeter, and the
negative lead to the negative terminal (Aout-).
2. Then in Meter Menu highlight [Analog Output (std)] and press [ENTER]. Then highlight [4-20mA] and
press [ENTER].
3. Scroll down and select [Calibrate] option.
Analog Output (Std)
Off
Measurement Type
Channel 1
Channel 2
Channel 3
Composite
Full Value
Full Value
Full Value
UNIT: ft/s
3.0000
3.0000 ft/s
Calibrate 4mA
UNIT: mA
4.000
Error Handling Value
Value
UNIT: mA
4.00
Calibrate
Calibrate 4mA
Calibrate 20mA
Percentage of Scale
Measurement Type/Comp
Measurement Type
4-20 mA
Calibrate
Base Value
Error Handling
Full Value
4-20 mA
Calibrate
Base Value
Error Handling
Full Value
4-20 mA
Calibrate
Base Value
Error Handling
Full Value
Input/Output
Analog Output (Std)
Digital Output (Std)
Option Comm (Slot 1)
Option I/O (Slot 2)
Error Handling
Low
High
Hold
Other
4-20mA
4-20mA
Base Value
Full Value
Calibrate
Measurement Type
Channel 2
Channel 3
Composite
Channel 1
Avg Vol Flow Rate
Soundspeed
Mass Flow
Measurement Type/Ch #1
Volumetric (Act)
Volumetric (Std)
Mass Flow
Velocity
Measurement Type/Ch #1
Soundspeed
Gain
Avg Vol Flow Rate
Std Deviation Gain
4.00 mA
Value
9
2
9
2
9
2
9
2
9
2
9
2
9
2
9
or
9
2
9
or
9
2
9
or
9
2
9
or
9
2
9
or
9
2
Volumetric (Act)
Volumetric (Std)
Mass Flow
Velocity
Measurement Type/Comp
Reynolds #
Measurement Type/Ch #1
Std Deviation Gain
SNR Up
Gain
SNR Dn
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
Operator Access or Higher

Table of Contents

Related product manuals