Chapter 3. Programming
58 PanaFlow™ LC User’s Manual
Figure 65: Setting up Option I/O Analog Input/RTD Input
3.5.5d Option IO (Slot2): Calibrating Analog Input
To calibrate Option I/O Analog Input, use steps as in section “Log-in and Primary Pages” to navigate to the 
Input/Output settings page.
Note:  Optional I/O menu uses Slot:Channel convention for clarity. For Example, Analog Input(S2:3) indicates 
Analog Input on Slot 2, Channel 3. Optional I/O is installed in the Slot2 of the electronics stack.
1. Connect the Analog Input as in Figure 66.
2. Turn ON the calibrator and set it to measure Current (mA) DC. Connect test lead from positive side 
(Analog Input Channel 3: I/O 7) of the optional I/O Analog Input to the positive terminal of 
multimeter, and the negative lead to the negative terminal (Analog Output Channel 3: I/O 8).
3. Highlight [Option I/O (Slot 2)] and press [ENTER].
4. Then highlight [AO-AO-AI-R-1000-3W] and press [ENTER]. Then highlight [Analog Input(S2:3)] and 
press [ENTER].
5. Then highlight [4-20mA] and press [ENTER].
6. Scroll down and select [Calibrate] option.
7. Select [Calibrate 4mA] in the meter Menu. Set [4mA] current on calibrated current source and check 
the reading on the XMT1000 is 4.00mA ±0.01mA. Once reading is stabilized on the LCD, Press [ENTER] 
to accept the current 4 mA value or Press [ESCAPE] to cancel the calibration.
8. Select [Calibrate 20mA] in the meter Menu. Set [20mA] current on calibrated current source and 
check the reading on the XMT1000 is 20.00mA ±0.01mA. Once reading is stabilized on the LCD, Press 
[ENTER] to accept the current 20 mA value or Press [ESCAPE] to cancel the calibration.
9. Once calibration is complete, select either [Save] or [Save & Logout] option to save calibration data.
Table 15: Analog Input Measurement Types
Measurement Options for Analog Input
Temperature, Pressure
Analog Input (S2:3)
Off
Measurement Type
Pressure
Temperature
Analog Input (S2:3)
Base Value
Full Value
Calibrate
Measurement Type
Set Point 1
UNIT:   C
0
Calibrate 4mA
Set Point 1
Set Point 1
0              C
4.000
UNIT:    mA
4-20mA Input
 
9 
 
2 
 
9 
 
2 
 
9 
 
2 
 
9 
 
2 
 
9 
 
9 
 
2 
 
9 
Input/Output
Option  I/O (Slot 2)
Analog Output (Std)
Digital Output (Std)
Option  I/O (Slot 2)
AO-AOAI-R-1000-3W
 
9 
 
2 
 
9 
 
2 
AO-AOAI-R-1000-3W
Analog Input (S2:3)
RTD Input (S2:4)
Analog Output (S2:1)
Analog Output (S2:2)
or
 
9 
 
9 
 
2 
Analog Input (S2:3)
Base Value
Full Value
Measurement Type
Calibrate
Calibrate
Calibrate 4mA
 
9 
 
2 
Calibrate 20mA
RTD Input (S2:4)
Off
Calibrate RTD
Read Value 1
Set Point 1
RTD Input (S2:4)
No of RTD Cal points
Calibrate
RTD Input
 
9 
 
2 
 
9 
 
2 
AO-AOAI-R-1000-3W
RTD Input (S2:4)
Analog Input (S2:3)
Analog Output (S2:1)
Analog Output (S2:2)
 
9 
 
2 
Set Point 2
Read Value 2
Calibrate RTD
Read Value 4
Set Point 5
Set Point 4
Read Value 5
[        ]
 
2
 
 
9 
[        ]
UNDO
MOVE
[        ]
[        ]
SAVE
ALTER
Operator Access or Higher
[        ]
 
2
 
 
9 
[        ]
UNDO
MOVE
[        ]
[        ]
SAVE
ALTER