PanaFlow™ LC User’s Manual 59
Chapter 3. Programming
Figure 66: Option I/O Analog Input connections
3.5.5e Option IO (Slot2): Calibrating RTD Input
1. Insert RTD sensor and master RTD in temperature bath and turn ON, and set it to desired
temperature set point.
2. Use steps as in section “Log-in and Primary Pages” to navigate to the Input/Output settings page.
Refer to Figure 65 above, highlight [Option I/O (Slot 2)] and press [ENTER].
3. Highlight [AO-AO-AI-R-1000-3W] and press [ENTER]. Then highlight [RTD Input(S2:4)] and press
[ENTER].
4. Highlight [RTD Input] and press [ENTER].
5. Set the number of Calibration points by selecting [No. of RTD Cal points].
6. Scroll down and select [Calibrate] option.
7. Select [Set point 1], press [ENTER] and set [Set point 1] to the temperature value selected on the
calibrator. Press [ESCAPE].
8. Select [Read Value 1], press [ENTER], and check the [Read Value 1] reads [Set point 1] value. Once
reading is stabilized on the [Read Value 1], Press [ENTER] to accept the value or Press [ESCAPE] to
cancel the calibration.
9. Repeat steps 7 and 8 for other set points.
10. Once all set points are calibrated, select either [Save] or [Save & Logout] option to save calibration
data.
Figure 67: Option I/O RTD Connection