4-32 250cx Series Maternal/Fetal Monitor Revision C
2036947-001
Maintenance: Testing the Tocotransducers
Uterine Activity Ramp
Testing the Tocotransducers
CAUTIONS
LEAKAGE TEST—Perform a leakage and dielectric test on the
tocotransducer per applicable standards.
FUNCTIONAL TEST—Perform a functional response test on the
tocotransducer. Inspect a Nautilus Tocotransducer as follows:
1. Check for any cracks or contaminants on the tocotransducer especially on the
diaphragm located on the bottom of the tocotransducer. Visibly inspect the
condition of the cable, strain relief, and connector pins.
2. Connect the tocotransducer to the UA input connector on the front panel of the
250cx Series Monitor.
IMPORTANT
TRIMLINE TOCOTRANSDUCER—If you are using an older Trimline
tocotransducer for this test, be advised of the following. If the monitor is on
when you connect or re-connect a Trimline Tocotransducer to the UA
connector, you must wait at least 10 seconds before pressing the UA
Reference button. If the monitor is off, you must wait at least 10 seconds
from the time the monitor is powered on.
3. Access the
Install Options Screen 2
and note the
Default TOCO Reference
setting. Exit
the service mode by selecting
Restart
at the bottom of the screen.
4. Momentarily depress the monitor’s UA Reference button. Verify the following: