Revision C 250cx Series Maternal/Fetal Monitor 3-21
2036947-001
Installation: Communications Setup Screen
Communications Setup Screen
The Communications Setup screen, allows you to configure the unit’s
J109, J110, and
J111 RS-232C
Serial Interface ports for connecting to optional peripheral equipment.
Each port is configured for baudrate and mode.
To use the Communications Setup screen:
Select the Service softkey from the General Setup screen.
The Service Lock screen displays.
Enter the correct access code; Install Options Screen 1 displays.
Select the COMM softkey on Install Options Screen 1.
The Communications Setup screen appears.
Baudrate
This field selects the baudrate for communication with an external device. Use the
Trim Knob control to cycle through the available settings: 600, 1200, 2400, 4800,
9600, and 19,200 bps. (Each port is factory set to 2400 bps.) Refer to “Peripheral
Components” on page 2-14 for more information about configuring this field.
Mode
This field selects the mode for communication with an external device. Use the Trim
Knob control to cycle through the available settings: Nellcor, Factory, Critikon, 1371,
1371/NOTES, 250Plus, LOOPBACK, 115, 115 X/R, and Exergen. (Each port is factory set
to the 1371/Notes mode.) Refer to “Peripheral Components” on page 2-14, for
more information about configuring this field.
NOTE: At the factory settings, the monitor is ready for connection to a Centricity
Perinatal system.