Revision C 250cx Series Maternal/Fetal Monitor 5-3
2036947-001
Troubleshooting: Diagnostic Tests
Diagnostic Tests
Main Motherboard and DSP Board Self-Test
Upon power-up, the 250cx Series Monitor’s Main Motherboard and DSP Board
perform a number of tests before allowing the monitor to enter the normal
operational mode. After setting the Power switch to the on (
I) position, the tests run
for approximately 2 seconds. If all tests pass, the monitor enters the normal
operational mode. During normal operational mode, the monitor continues to
perform a series of self-tests. If a test fails, you will see one of the following error
messages centered in the display:
SYSTEM FAULT: ROM
SYSTEM FAULT: RAM
SYSTEM FAULT: Alert
SYSTEM FAULT: UI KEYPAD
SYSTEM FAULT: SOFTWARE
SYSTEM FAULT: DEFAULTS
SYSTEM FAULT: DSP
The first errors indicate problems on the Main Motherboard. The last error indicates
a problem on the DSP Board. If the monitor displays any of these messages, turn off
the monitor and contact your GE Service Representative.
Monitor Self-Test
The 250cx Series Monitor contains self-test routines which verify the unit’s
calibration and internal circuitry. These routines are initiated by depressing the front
panel Test button. The test results are printed on the strip chart recorder paper,
verifying the integrity of the unit.
It is recommended practice to initiate the self-test feature at the beginning of each
monitoring session. Refer to Self-Test Routine in the “Maintenance Section.”
Error Log Screen
The Error Log screen, as shown below, displays a service log of the 250cx Series
Monitor. The service log is useful only to GE engineering and may be requested by
your service representative. There are 255 error codes that can be detected by the
monitor. The Error Log is comprised of multiple pages with up to 20 error codes per
screen. Each screen displays three columns:
The Error ID column lists the error code. The Count column lists the number
times the error code has occurred.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals