Revision C 250cx Series Maternal/Fetal Monitor 4-53
2036947-001
Maintenance: Recorder Photosensor Calibration
Adjusting the Paper-Loading Sensor
1. Load paper in the recorder. Ensure that there are
no
black squares showing to
indicate a paper-low condition. You must load at least nine sheets of paper.
2. Using a digital voltmeter, connect the positive lead to J9 (pin 6) on the Recorder
Board; connect the negative lead to J9 (pin 2).
3. Adjust R41 for a reading of +190 mVdc ± 5 mVdc on the digital voltmeter.
4. Although a factory-only fixture is required for this step, you may wish to try it
by estimating the required distance. Load paper backwards with the white
surface 0.380 in away from the paper tray. This creates a paper-loading error
condition. Verify that the voltage at J9 (pin 6) on the Recorder Board is at least
2.70 Vdc.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals