6-12 250cx Series Maternal/Fetal Monitor Revision C
2036947-001
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
2025177-009 Isolated Power Supply Board
1. Remove top cover. Keep screws aside.
2. Remove two screws fastening DSP board to chassis.
3. Remove DSP board from board.
4. Disconnect 8-conductor Inverter cable from DSP board connector J1.
5. Disconnect 20-conductor DSP-to-LCD Decoder ribbon cable from DSP board
connector J2.
6. Remove card cage cover. Keep screws aside.
7. Remove Isolated Power Supply board from Front-end Motherboard.
8. Insert new Isolated Power Supply board into Front-End Motherboard.
9. Replace card cage cover. Re-insert screws.
10. Reconnect 8-conductor Inverter cable to DSP board connector J1.
11. Reconnect 20-conductor DSP-to-LCD Decoder ribbon cable to DSP board
connector J2.
12. Place DSP board onto board. Use two screws to fasten DSP board to chassis.
13. Place top cover. Re-insert screws.
2025177-010 SpO
2
Carrier Board with Nellcor MSpO
2
Module
1. Remove top cover. Keep screws aside.
2. Remove two screws fastening DSP board to chassis.
3. Remove DSP board from Main board.
4. Disconnect 8-conductor Inverter cable from DSP board connector J1.
5. Disconnect 20-conductor DSP-to-LCD Decoder ribbon cable from DSP board
connector J2.
6. Remove card cage cover. Keep screws aside.
7. Disconnect MSpO
2
cable from SpO
2
Carrier board connector.
8. Remove SpO
2
Carrier board with Nellcor MSpO
2
Module assembly from
Front-end Motherboard.
9. Insert new SpO
2
Carrier board with Nellcor MSpO
2
Module assembly into
Front-End Motherboard.
10. Reconnect MSpO
2
cable to SpO
2
Carrier board connector J10.
11. Replace card cage cover. Re-insert screws.
12. Reconnect 8-conductor Inverter cable to DSP board connector J1.
13. Reconnect 20-conductor DSP-to-LCD Decoder ribbon cable to DSP board
connector J2.
14. Place DSP board onto Main board. Use two screws to fasten DSP board to
chassis.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD