Revision C 250cx Series Maternal/Fetal Monitor 6-13
2036947-001
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
15. Place top cover. Re-insert screws.
2025177-011 SpO
2
Carrier Board with Masimo MSpO
2
Module
1. Remove top cover. Keep screws aside.
2. Remove two screws fastening DSP board to chassis.
3. Remove DSP board from Main board.
4. Disconnect 8-conductor Inverter cable from DSP board connector J1.
5. Disconnect 20-conductor DSP-to-LCD Decoder ribbon cable from DSP board
connector J2.
6. Remove card cage cover. Keep screws aside.
7. Disconnect MSpO
2
cable from SpO
2
Carrier board connector.
8. Remove SpO
2
Carrier board with Masimo MSpO
2
Module assembly from
Front-end Motherboard.
9. Insert new SpO
2
Carrier board with Masimo MSpO
2
Module assembly into
Front-End Motherboard.
10. Reconnect MSpO
2
cable to SpO
2
Carrier board connector J10.
11. Replace card cage cover. Re-insert screws.
12. Reconnect 8-conductor Inverter cable to DSP board connector J1.
13. Reconnect 20-conductor DSP-to-LCD Decoder ribbon cable to DSP board
connector J2.
14. Place DSP board onto Main board. Use two screws to fasten DSP board to
chassis.
15. Place top cover. Re-insert screws.
2025177-012 SpO
2
Carrier Board with TruSignal MSpO
2
Module
1. Remove top cover. Keep screws aside.
2. Remove two screws fastening DSP board to chassis.
3. Remove DSP board from Main board.
4. Disconnect 8-conductor Inverter cable from DSP board connector J1.
5. Disconnect 20-conductor DSP-to-LCD Decoder ribbon cable from DSP board
connector J2.
6. Remove card cage cover. Keep screws aside.
7. Disconnect MSpO
2
cable from SpO
2
Carrier board connector.
8. Remove SpO
2
Carrier board with TruSignal MSpO
2
Module assembly from
Front-end Motherboard.
9. Insert new SpO
2
Carrier board with TruSignal MSpO
2
Module assembly into
Front-End Motherboard.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD