4-50 250cx Series Maternal/Fetal Monitor Revision C
2036947-001
Maintenance: Calibration
Calibration
This section of the manual provides a calibration procedure which allows authorized
service personnel to perform an instrument alignment using a minimum of test
equipment. This procedure is not intended to replace a complete instrument
checkout and alignment as performed at the GE factory. It should be considered a
performance check and troubleshooting guide to be used in conjunction with other
information supplied throughout this service manual. It is important to mention, this
section of the manual is not intended as a substitute for proper professional training,
or familiarity with the monitor. Only qualified service personnel should attempt
servicing the 250cx Series Monitor.
Before You Begin Electronic Calibration
General
Refer to the assembly drawings for the location of adjustments and test points
referred to in this section. The board assembly drawing numbers are given for each
board in this section. The electronic calibration procedure outlined in this section
necessitates removal of the monitor’s top cover. Also, the following equipment will
be needed in order to carry out the procedural steps.
Digital Voltmeter, 4 1/2 digit or equivalent
Plastic Alignment Tool
Oscilloscope
Handling Precautions
The following guidelines should be followed when handling circuit boards
or assemblies containing circuit boards. Following these procedures helps
resist damage that can be caused by static electricity.
Discharge any static charge you may have built up before handling parts.
Wear a grounded, anti-static wristband at all times.
Use a static-free work surface.
Store items in anti-static bags or boxes.
Do not remove items from anti-static containers until needed.
Power Supply Voltages—Verification
There are no adjustments for the power supply voltages; however, you can verify
them as follows:
Main Board Power Supply Voltages
Using a digital voltmeter, measure and verify the power supply voltages using the
J14 connector on the Main Board. Refer to Table 13.
NOTE: There are no adjustments for these voltages.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals