Revision C 250cx Series Maternal/Fetal Monitor 5-7
2036947-001
Troubleshooting: Diagnostic Control Screen
Run Time
This field displays the amount of time the monitor has been turned on—since the
field was cleared. The time is displayed in hours, minutes, and seconds. To clear this
field (reset the timer), activate the Clear softkey to the right of the run time field.
NOTE: The Clear softkey also resets the recorder time field.
Recorder Time
This field displays the amount of time the recorder has been turned on (or printing in
maternal-only mode)—since the field was cleared. The time is displayed in hours,
minutes, and seconds. To clear this field (reset the timer), activate the Clear softkey
to the right of the recorder time field.
NOTE: The Clear softkey also resets the monitor run time field.
Recorder Servicing
Removing the Strip Chart Recorder
Removing the strip chart recorder requires loosening the monitor’s front panel bezel.
Following this procedure carefully will avoid unnecessary damage to bezel’s
mounting screws and the unit’s front panel connectors.
1. Turn off the 250cx Series Monitor and disconnect the power cord from the
monitor.
2. Remove the nine screws which secure the monitor top cover. Four screws are
located on the bottom of the monitor and five screws are located in the back.
3. Remove the cover by sliding it toward the rear of the monitor.
4. Remove the four screws that attach the recorder bracket to the bottom of the
chassis.
5. Remove the two screws which fasten the right and left sides of the bezel away
to the chassis.
6. Disconnect the display ribbon cable from the DSP Board at J5.
7. Disconnect the ribbon cable from the DSP Board.
8. Disconnect the ribbon cable from the Recorder Board at J2. Leave the other end
connected to the Main Motherboard.
9. Disconnect the Power Supply cable harness from the Recorder Board at J1.
10. Lift the recorder and its attached bracket out of the monitor.
Installing the Strip Chart Recorder
1. Carefully place the recorder into the monitor housing and reassemble to the
chassis bottom with four screws.
2. Re-connect the Power Supply cable harness to the Recorder Board at J1.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power