4-52 250cx Series Maternal/Fetal Monitor Revision C
2036947-001
Maintenance: Recorder Photosensor Calibration
Recorder Photosensor Calibration
Adjusting the Paper-Low Photosensor
1. Load paper in the recorder. Ensure that there are
no
black squares showing to
indicate a paper-low condition.
2. Press the Record button to turn on the recorder. Allow the paper to advance for
a few seconds in order to tension the paper.
3. Turn off the recorder.
4. Using a digital voltmeter, connect the positive lead to J9 (pin 4) on the Recorder
Board; connect the negative lead to J9 (pin 2).
5. Adjust R31 for a reading of +150 mVdc ± 2 mVdc on the digital voltmeter.
NOTE: If you open and then close the recorder door, the reading may vary 5–10
mV, due to the loss of tension in the paper. This is acceptable and you do
not need to re-adjust.
6. Open the door and verify that the reading on the digital voltmeter is greater than
+4.75 Vdc.
7. Re-load the paper so that black squares show on the surface (i.e., the last several
sheets of a pack).
8. Turn on the recorder.
9. The value on the digital voltmeter should go up and down as the paper surface
alternates between black and white. Verify that the maximum value is greater
than or equal to 2.0 Vdc.
10. Turn off the recorder.
Adjusting the Paper-Out Photosensor
1. Load paper in the recorder. Ensure that there are
no
black squares showing to
indicate a paper-low condition.
2. Press the Record button to turn on the recorder. Allow the paper to advance for
a few seconds in order to tension the paper.
3. Turn off the recorder.
4. Using a digital voltmeter, connect the positive lead to J9 (pin 3) on the Recorder
Board; connect the negative lead to J9 (pin 2).
5. Adjust R29 for a reading of +150 mVdc ± 2 mVdc on the digital voltmeter.
NOTE: If you open and then close the recorder door, the reading may vary 5–10
mV, due to the loss of tension in the paper. This is acceptable and you do
not need to re-adjust.
6. Open the door and verify that the reading on the digital voltmeter is greater than
+4.75 Vdc.
7. Turn off the recorder.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals