5-4 250cx Series Maternal/Fetal Monitor Revision C
2036947-001
Troubleshooting: Error Log Screen
The Data column lists other information associated with the error code. This
column is for factory use only.
Error Log Screen
To use the Error Log screen:
1. Select the Service softkey from the General Setup screen.
2. The Service Lock screen appears.
3. Enter the correct access code; the Install Options Screen 1 displays.
a. Select NextPage.
b. The Install Options Screen 2 displays.
4. Select the Log softkey from the Install Options Screen 2.
5. The first Event Log screen displays.
6. Use the Page softkey to display the next screen (page), if applicable. The last
page wraps back to the first page.
7. Use the Print softkey to print the displayed screen (page) on the strip chart
recorder.
8. Use the Clear softkey to clear all the error log from the battery-backed RAM.
Error Log
Error LogCountData
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
Print
Clear
Page
Exit

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals