Revision C 250cx Series Maternal/Fetal Monitor 4-57
2036947-001
Maintenance: Preventative Maintenance Inspection Report
Parameter Cond. 1
Required
(Normal)
Cond. 2 Cond. 3 Cond. 4 Cond. 5 Cond.6
Required
(Single Fault)
Chassis <100 µA <500 µA
MECG Ld/Gnd <10 µA <50 µA
MECG Ld/Line <10 µA <50 µA
FECG Ld/Gnd <10 µA <50 µA
FECG Ld/Line <10 µA <50 µA
IUP/Toco Ld/Gnd <10 µA <50 µA
IUP/Toco Ld/Line <10 µA <50 µA
US1 Ld/Gnd <10 µA <50 µA
US1 Ld/Line <10 µA <50 µA
US2 Ld/Gnd <10 µA <50 µA
US2 Ld/Line <10 µA <50 µA
SpO
2
Ld/Gnd <10 µA <50 µA
SpO
2
Ld/Line <10 µA <50 µA
Functional Checkout Tests
Page
Description
Complete Max Actual Pass Fail N/A
5-3 Monitor Self-Test
4-19 Front Panel Button Test
4-20 MECG Test
4-23 FECG Test
Leg Plate Test (Follow Manufacturer’s Directions)
4-26 Ultrasound Test
4-28 Fetal Movement Detection Test
4-29 Ultrasound Test
4-30 Uterine Activity Test
4-32 Tocotransducer Test
4-33 Strain Gauge Test (if used)