Revision C 250cx Series Maternal/Fetal Monitor 5-15
2036947-001
Troubleshooting: Diagnostic Control Screen
4. Install a new cable tie to hold the cable in place.
NOTE: Ensure that there is enough slack in the cable so that when the recorder
door is opened the cable does not become taut.
Re-install the strip chart recorder. Refer to “Installing the Strip Chart
Recorder” on page 5-7. Do not replace the monitor top cover yet.
5. Adjust the paper-loading sensor as described next in this section.
Adjusting the Paper-Loading Sensor
Refer to “Recorder Photosensor Calibration” on page 4-52 for detailed
instructions.