B-4 250cx Series Maternal/Fetal Monitor Revision C
2036947-001
Alarms Summary:
NOTE: For further information on Alarms, refer to Chapter 11 in the
Corometrics 250cx Series Operator’s Manual.
MHR/P
1
Alarm Defaults
Audio: on
Volume: 5
Limits: High = 120 bpm, Low = 50
bpm
Re-alarm: 120 sec
—
MHR/P limit (high or low) actively
being violated.
The tachycardia response time is < 8
seconds.
MHR/P numeric flashes. Alternating high/low tones (if
audio enabled).
Asystole. Flashing dashes “– – –” in place
of MHR/P numeric.
Alternating high/low tones (if
audio enabled).
MECG leads off. Flashing dashes “– – –” in place
of MHR/P numeric and MECG
LEADS OFF message displays
underneath.
Alternating high/low tones (if
audio enabled).
1
There is an MECG re-alarm.
MSpO
2
Alarm Defaults
Audio: on
Volume: 5
Limits: High = 100%, Low = 95%
Re-alarm: 120 sec
—
MSpO
2
limit (high or low) actively
being violated. Issued after about 8
seconds.
MSpO
2
numeric flashes. MSpO
2
value and pulse rate print on the
strip chart.
Alternating high/low tones (if
audio enabled).
Malfunction with MSpO
2
circuitry. COMM or REPAIR message
displays in MSpO
2
area.
Alternating high/low tones (if
audio enabled).
MSpO
2
intermediate cable
disconnected from monitor; sensor
assembly disconnected from
intermediate cable; or sensor or
cable has a broken wire.
Dashes “– – –” in place of MSpO
2
numeric.
Alternating high/low tones (if
audio enabled).
Table B-1. Summary of 250cx Series Alarms
Type Condition Visible Advisory Audible Advisory