Revision C 250cx Series Maternal/Fetal Monitor 2-19
2036947-001
Equipment Overview: Theory of Operation
Video Flex
data cable
Main Board/NIBP
The main board houses the central processor and is the
focal point for all data in the 120. The board is responsible
for the following:
Receives processed FECG, Ultrasound, UA, and
MECG data from the DSP board
Formats DSP data for display
Passes display data to the DSP board display
FPGA which in turns drives the LCD panel.
Writes the processed data to the recorder board
Receives front panel keypad/trimknob data from
the UI board
Communicates parameter data to the rear panel
comm. Ports
Passes parameter telemetry data to the DSP board
fro processing.
Inputs keypad data from rear panel comm. Board
and sends it to the recorder board
Executes diagnostics
o Reports its own main failures
o Reports UI keypad and DSP failures
o Recorder test pattern and display test
pattern
o
Audio test pattern
Provides NIBP processing subsystem &
communicates to the external Pneumatics
PWA.
DSP/Display Interface Board
The DSP/Display interface board forms the
bridge between the front–end parameters and
the main board central processor. It also has
the function of display interface. It interfaces
the Main board to the LCD panel by
providing a dual port memory to the main
board and a graphics engine driving the LCD
panel. The board is responsible for the
following:
Interfaces to front-end motherboard
which holds the parameter boards
DSP Processing of FECG, MECG,
UA, and Dual Ultrasound data thus
converting analog data to heart rates
and pressure values.
Sending of processed data to main
board
Pass through of MSpO2 OEM
module serial communication lines to
the main board (no processing done
by DSP)
Provides status of patient cables
plugged in to monitor to the main
board
Drives the LCD display panel by
generating the video graphics through
the FPGA chip and sending data to
the decoder board which then
connects to the LCD display
Interfaces the video to the main board
processor through a dual port memory
Pass through of front panel serial
comm lines to the main board
Generates VGA video output to rear
p
anel
Digitally Processed
Analog Parameter
Data
Un-Processed Analog
Parameter Data
From telemetry
Graphics Data
for display
Pass through
MSpO2 & FSpO2
serial data & Reset
Recorder Board
Interfaces main board serial data to recorder
printhead
Provides regulated power to printhead
Provides sensor data from recorder door, paper
out & paper misload to ma
in board
Pass through UI
Keypad serial data
Front-end
Motherboard
Connects front-
end bds together
distributing
isolated power
and un-isolated
power.
Passes through
all analog signals
to the DSP board
for processing.
FECG/UA Board
Converts fetal ECG to analog
signal for processing
Converts pressure signal to
analog for DSP signal
processing
MECG Board
Converts 3 lead maternal ECG
signals to analog signal for
DSP processing
Dual Ultrasound Board
Provides doppler shift analog
signals to DSP board
SpO2 Carrier Board
Carries MspO2 module
connecting to FE mthbd
MspO2 Module
Provides serial MspO2 data to
main board
Isolated Power Supply
Provides isolated power to all
front-end boards excluding US
board
All
Analog
Signals
Telemetry
US data
UI Keypad
Passes all button closures and trimknob data to
main bd via DSP board.
Provides the chart light & key backlight function
under main board control
Interfaces serially to the main bd
Serial keypad
data cab
le
Comm Board
Interfaces ext keypad to main board
Interfaces ext telemetry analog signals to main bd
Interfaces legacy Spectra analog central by
sending
.
Telemetry
analog signals
Keypad
signals
Analog HR
outputs
Analog HR signal
with Mode Info
Analog HR signal
with Mode In
fo
Lead selection,
Pacer enable, test
signal enable
Serial digital MpO2/
data
MSpO2 Resets
Serial digital MpO2
data
FECG +
MECG
Front panel
Connector
FECG UA cables (2)
MECG
cable
Analog HR signal
with Mode In
fo
FECG & UA signals
Power for
MspO2/FspO2,
FECG, and MECG
b
oards
System +20 volts
for switching
l
Serial digital
MpO2/FSpo2
data
MSpO2/
FSpO2
Resets
UI Volume
Keypad
Holds volume
& alarm
cancel buttons
US1
Connector
US2
Connecto
r
MspO2 FP
Connector
Bd
UA Front
Panel
Connector
UA cable
US1 cable
US2 cable
MS
p
O2 cable
Isolated
Isolated
Denotes Patient
Isolated Sections
Video Decoder
Board
Translates 4 bit
color to 18 bit
color for LCD
LCD
Panel
320 X 240
Inverter
Module
Volume
data cable
DSP/Decoder
Video cable
DSP Inverter
Power cable
LCD Inverter
cable
(part of LCD)
Recorder Serial
data/status
Cable
System Power
Supply
+5, +12, +15, -15
To Main
Boar
d
To Recorder
Boar
d
Power Cables
Power Cable from
system power
supply
Power Cable from
system power
supply
Pneumatics
Board
Contains
pump, valves
and filte
r
Denotes
Plug-in
connections
Block Diagram Description
Isolated
Isolated
Isolated
Isolated
Isolated
Partially
Isolated
Denotes cables
Isolated
VGA Output
Standard VGA
output through
15 pin HD conn
New VGA
Video cable
to rear panel
New or Changed
for Cheetah
Temp.
Module
External
Temperature
Module

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD