Revision C 250cx Series Maternal/Fetal Monitor 2-25
2036947-001
Equipment Overview: Theory of Operation
FECG signals as well as the MECG mode lines. The MECG analog signals are
routed separately from the front panel to the MECG daughter PWA. The MSpO
2
input signals enter the SpO
2
carrier PWA directly.
Video Decoder PWA
The decoder PWA interfaces between the DSP/Display PWA and the LCD panel.
The decoder PWA performs conversion of 4-bit color information from the FPGA
output to the 18-bit color required by the LCD panel. This provides a 16-color
palette. The translation is accomplished in the FPGA on the decoder PWA. This
PWA receives high-speed video from the DSP/Display PWA.
Recorder PWA
The recorder PWA is responsible for driving the recorder motor and the recorder
printhead device along with providing the main system PWA with paper out/low/
misload status. To drive the motor it receives pulses from the main system PWA and
provides the proper drive circuitry to drive the stepper motor 4 phase windings. To
drive the printhead an adjustable power supply is provided which is set to the
printhead specifications (each printhead is unique). Data to be printed and control
information is received from the main system PWA, buffered and presented to the
printhead. Sensors from the recorder assembly to detect paper low/out/misload are
received and translated to digital status lines to be sent to the main system PWA.
Communications PWA
The Communications PWA contains three basic interfaces. It supports the analog
interface (J102) to the legacy Spectra 400 surveillance system as well as other
manufacturers centrals, a 2116 external keyboard interface for strip chart annotation,
and an analog telemetry interface to the 340 telemetry system. The communications
PWA communicates to the system by directly plugging into the main system PWA.
Digital data from the keypad interface and telemetry (modes only) is transferred
through a data bus and analog signals (MECG, FECG, Ultrasound, TOCO) from the
telemetry are separately routed through the main PWA to the DSP for processing
similarly to the existing front panel patient data.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals