Revision C 250cx Series Maternal/Fetal Monitor 3-15
2036947-001
Installation: Service Mode Screens
Language
This field is used to set the language shown on the display and printed on the strip chart
recorder paper. Use the Trim Knob control to cycle through the available options:
English, Spanish, French, German, Dutch, Swedish,
Italian, Danish, Finnish, and Norwegian
.
(The monitor is factory set for the country in which it is being used.)
Line Frequency
This field is used to define the line power frequency for the country in which the
monitor is being used. Use the Trim Knob control to alternate between 50 Hz and 60
Hz. (The monitor is factory set for the country in which it is being used.)
Scaling
This field is used to set the heart rate scale for the strip chart paper. Use the Trim
Knob control to alternate between 30–240 bpm/cm and 50–210 bpm/cm. (The monitor
is factory set for the country in which it is being used.) The MSpO
2
trend Auto scale
is also affected by this setting. When the heart rate scale is 30–240, the MSpO
2
expanded scale is 60–100% (10%/cm). When the heart rate scale is 50–210, the
MSpO
2
expanded scale is 50–100% (12.5%/cm).
Recorder Font Size
The monitor offers a choice of font sizes to print annotations.
Small: increases print speed.
Medium: a compromise between the large and small sizes.
Large: fosters readability.
VS Print Interval
The monitor provides an option for the printing of blood pressure and MSpO
2
values
on the strip chart paper:
Real-Time: Values are printed according to the actual clock time (9:33, 9:48,
10:03, etc.).
Chart-Style: Values are printed on standard clock quarter (9:00. 9:15, 9:30:, 9:45,
etc), half (9:00, 9:30, 10:00, 10:30, etc), and whole hour marks (9:00, 10:00,
11:00, 12:00, etc).
Paper Chime
This field enables/disables an audible tone to indicate a low-paper or out-of-paper
condition.
Off: no audio tone sounds with either condition; however, the visual indicator is
still present (the recorder LED blinks).
Low/Out: sounds a tone when either LOW BATTERY or OUT OF PAPER condition
exists.
Out Only: sounds a tone only when the OUT OF PAPER condition exists.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD