Revision C 250cx Series Maternal/Fetal Monitor 3-19
2036947-001
Installation: Service Mode Screens
NIBP 1 min Interval
This field is used to enable/disable the 1 minute interval selection on the maternal
BP Setup screen. Use the Trim Knob control to alternate between On and Off. (The
factory default setting is Off.)
NIBP Display
Setting this field to a value reduces the chance of error. Setting this field to On leaves
the blood pressure reading displayed indefinitely and could potentially cause
confusion. For example: if the monitor is configured for manual mode and 1 hour
has elapsed since the last reading, the continuous display of the “old” NIBP reading
may cause confusion.
This field determines the time period, in minutes, that a blood pressure reading
remains displayed before being automatically erased,
1
starting from the time the
reading is displayed.
Pressure Units
The Pressure Units allows you choose either mmHg or kPa.
SatSeconds
The SatSeconds limit controls the time that the SpO
2
level may fall outside the alarm
before an audible alarm sounds. The SatSeconds default setting is set at 10
SatSeconds. Your biomed may set the SatSeconds limit (Off, 10, 25, 50, or 100) to suit
the clinical environment and patient condition.
Default Settings
The 250cx Series Monitor is shipped with factory defaults for the setup screens. You
can change most of these settings to suit your particular needs. The monitor has two
Default Settings: that allow you to determine the settings on the next power-on or
restart: Factory or Hospital.
Factory: Select this option if you wish to return all monitor settings back to the
factory default settings (refer to the table at the end of this section). Factory
defaults represent settings chosen to meet the requirements of a typical labor
and delivery environment.
NOTE:You may adjust the setup screen configurations as needed during
monitoring; however, be advised that if you turn off the monitor, all
user setup screens revert to the factory defaults when the monitor is
turned on again.
Hospital: The Hospital option allows your hospital to configure its own set of
preferences to be used each time the monitor is powered on. Record your
hospital settings in the table at the end of this section. You may adjust the setup
screen configurations as needed during monitoring; however, be advised that if
you turn off the monitor, all user setup screens revert to the factory defaults
when the monitor is turned on again.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals