3-26 250cx Series Maternal/Fetal Monitor Revision C
2036947-001
Installation: Factory Defaults
Install
Options
Screen 1
(Service)
Language Set according to shipping
destination
Set according to shipping
destination
Line Frequency United States: 60 Hz
International: 50 Hz
50 Hz, 60 Hz
Scaling United States: 30–240 bpm
International: 50–210 bpm
United States: 30–240 bpm
International: 50–210 bpm
Recorder Font Size Medium Small, Medium, Large
FECG Artifact Elimination Off On, Off
Paper Chime Out only Off, Low/out, Out only
Paper Chime Volume 5 1-9
HBC (Heartbeat
Coincidence Enable)
On On, Off
HR Offset
(Applies to US or US2—
whichever is FHR2)
10 min Off, On, 10 min
FM (Fetal Movement)
Remote Marker
On On, Off
SpO
2
Scale 0–100% Auto, 0-100%
(Does not change)
Install
Options
Screen 2
(Service)
Fetal Alert/Alarm Off Off, Alarms, Alerts
Alert Suspend Off Off, On
Re-Alarm
(MECG and SpO
2
only)
120 sec 120 - 300 seconds in 5-second
intervals
VS (Vital Signs) Print
Interval
Real Time Real Time, Chart Style
Default TOCO Reference 10 in mmHg mode or 1.3 in
kPa modes
5, 10, 15, 20, or 25 relative units
in mmHg mode or 0.7, 1.3, 2.0,
2.7, or in 3.3 kPa mode
Smart BP On On, Off
NIBP 1-min Interval On On, Off
NIBP Display On On, 1, 2, 3, 5, 10, 15, 30 min
Pressure Units mmHg
China: kPa
mmHg, kPa
SatSeconds (Nellcor) 10 Off, 10, 25, 50, 100
Default Settings Factory Factory, Hospital
Table 1. Summary of Factory Defaults
Setup
Screen
Field Description Factory Default Default Options Hospital Preference