Page 30 of 79 [EN] English – K0571 Issue 1
User Manual Druck DPI611
Table 2-1
Step Procedure
Use an applicable seal for the p
i. NPT type: Use an applicable sealant on the thread.
ii. BSP (parallel) type: We recommend a bonded seal at
the bottom.
Attach the adaptor to the device; if necessary use one of
the alternative adaptors listed in Section 1.11
(Accessories), then tighten to the applicable torque.
attach the adaptor to the test port and tighten it until
it is hand tight only.
b. Procedure (to Remove)
To remove a device, first release the pressure (See
Section 2.1.2). You can then do steps 4, 3, and 1
mentioned in Section 2.1.3 but do the operations in
the reverse order.
2.1.4 Vacuum or Pressure Operation
1 2 3 4 5
Figure 2-5 Vacuum or Pressure operation