Page 70 of 79 [EN] English – K0571 Issue 1
User Manual Druck DPI611
3. Select range (if applicable)
4. Follow on-screen instructions
When the calibration is complete, set the next calibration date.
5.5 PROCEDURES: CURRENT (measure)
Do the procedure as follows:
1. Connect the applicable calibration equipment (See Table 5-1)
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (See Section 5.4) to do a three-point
calibration (-FS, Zero and +FS) for each range: 20 mA and 55
mA.
4. Check the calibration is correct.
Select the applicable Current (measure) function.
Apply the values that follow:
mA: -55, -25, -20, -10, -5, 0 (open circuit)
mA: 0, 5, 10, 20, 25, 55.
Check the error is in the specified limits (See Table
5-2).
Table 5-2 Current (measure) error limits
Applied
mA
Calibrator
uncertainty
(mA)
Permitted
DPI611 error (mA)
±55 0.0030 0.0057
±25 0.0025 0.0043
±20 0.00063 0.0022
±10 0.00036 0.0016
±5 0.00025 0.0013
0 (open circuit) 0.0002 0.0010