Page 66 of 79 [EN] English – K0571 Issue 1
User Manual Druck DPI611
4.4.1 Sequence to Upload and Download file
Table 4-1
Step Procedure
1
Connect
the
USB
cable
(
See
Section
1.11
,
Accessories
)
to the Druck DPI611 calibrator.
2
Connect the lead to the USB port on the computer
that has the calibrator manager installed.
3
Use 4 Sight to set
-
up the procedure and create a
work order for the device.
The procedure includes the parameters for the
calibration, the number of test points, the relationship
and pass/fail tolerance.
4
Us
e the Download button in calibrator manager to
Download the file to the Druck DPI611 calibrator. A
communications symbol will be displayed at the
bottom of the screen.
5
Select:
DASHBOARD >> DOCUMENTING >> RUN PROCEDURE
6
In the Results window, select t
he filename specified in
calibrator manager.
7
Enter the User ID and the DUT Serial Number.
8
Tap on the Start button . The procedure sets up the
necessary Channel options e.g. mA and Volts.
9
Use the Take Reading button at each point specified
by the procedure. A prompt show for each point.
10
When all the readings are complete, tap on the Exit
Button . Look at the results on the display
(As found/As Left).
11
To complete the process, use the calibrator manager
to Upload the file back into the 4 Sight database.