Page 74 of 79 [EN] English – K0571 Issue 1
User Manual Druck DPI611
5.8 PROCEDURES: DC Volts (source)
Do the procedure as follows:
1. Connect the applicable calibration equipment (See Table 5-1).
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (See Section 5.4) to do a two-point
calibration for the applicable range:
Volts (source) range
10 V
24 V
4. Check the calibration is correct:
• Select the applicable Voltage (10V) or Current (24V)
(measure) function (See Section 5.7).
10V
24V
• Check the error is within limits (See Table 5-6).
Table 5-6 Voltage (source) error limits
Source
V