Page 73 of 79 [EN] English – K0571 Issue 1
User Manual Druck DPI611
Volts (V): 0, 5, 10, 20, 21, 30.
Check that the error is within limits (See Table 5-4
and Table 5-5).
Table 5-4 Millivolts (measure) error limits
Table 5-5 Voltage (measure) error limits
Applied
mV
Calibrator
uncertainty
(mV)
Permitted
DPI611 error
(mV)
±2000 0.051 0.1280
±1000 0.040 0.0940
±200 0.051 0.0148
±100 0.0040 0.0110
0 (short circuit) 0.0036 0.0070
Applied
V
Calibrator
uncertainty
(V)
Permitted
DPI611 error
(V)
±30 0.00052 0.00180
±21 0.00040 0.00154
±20 0.00031 0.00118
±10 0.00016 0.00089
±5 0.00008 0.00075
0 0.000024 0.00060