Page 72 of 79 [EN] English – K0571 Issue 1
User Manual Druck DPI611
Table 5-3 Current (source) error limits
5.7 PROCEDURES: DC mV/Volts (measure)
Do the procedure as follows:
1. Connect the applicable calibration equipment (See Table 5-1).
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (See Section 5.4) to do a three-point
calibration.
(-FS, Zero and +FS) for the applicable set of ranges:
mV (measure) ranges Volts (measure) ranges
200 mV 20 V
2000 mV 30 V
4. Check the calibration is correct:
Select the applicable Millivolts or Voltage (measure)
function.
Apply the input values that are applicable to the
Calibration:
mV: -2000, -1000, -200, -100, 0 (short circuit)
mV: 0, 100, 200, 1000, 2000
Volts (V): -30, -21, -20, -10, -5, 0 (short circuit)
Applied
mA
Calibrator
uncertainty
(mA)
Permitted
DPI611 error
(mA)
0.2 0.00008 0.0013
6 0.00023 0.0017
12 0.00044 0.0020
18 0.0065 0.0024
24 0.0012 0.0028