Page 69 of 79 [EN] English – K0571 Issue 1
User Manual Druck DPI611
PERFORM CALIBRATION
Table 5-1
Before starting the calibration, check the time and date on the
instrument are correct.
To do a calibration on a measure or source function, use the
advanced menu option.
Select:
DASHBOARD >> SETTINGS
>> ADVANCED
Enter the calibration PIN: 4321
Select the button.
Select:
Then select a function and start the calibration:
1. Select channel
2. Select function
Function
Calibration equipment
(ppm = parts per million)
Current
Current (mA) calibrator.
Accuracy - Current measure/source,
See Table 5-2 and Table 5-3.
Voltage
Volts calibrator.
Accuracy - Voltage measure/source,
See
Table 5-5 or Table 5-6.
Millivolts
mV calibrat
or.
Accuracy - Millivolts measure/source,
See Table 5-4 .
Pressure
(P1)
All Ranges total uncertainty of 0.025%
reading or
better.
IDOS
UPM only. Refer to the user manual for the IDOS UPM
.