After the Exam is Over
3-60
LOGIQ E9
User Guide
Direction 5454884-100 English
Rev. 1
Operator Controls
Cleaning Agents
ONLY use the following cleaners on the LOGIQ E9 Operator
Panel:
Palmolive Dishwashing Liquid (manufactured by
Colgate-Palmolive)
Sani Wipes (Sani-Cloth HB, green cap) Alcohol-free
(manufactured by Micorgen Inc.)
T-Spray II (manufactured by Pharmaceutical Innovations,
Inc.)
• Ammonia
Bleach (10 to 1 ratio of 5% home bleach)
Hydrogen Peroxide / Hydrogen Peroxide Wipes
DO NOT USE:
T-Spray I (Original T-Spray)
SaniCloth Plus (red cap)
Any cleaning/disinfecting solution containing Alcohol.
NOTE: When cleaning the operator control panel, make sure not to spill
or spray any liquid on the controls, into the system cabinet, or in
the probe connection receptacle.
NOTE: In case of SARS, use bleach or Cidex in a normal diluted form
for cleaning/disinfecting the operator panel.