EasyManua.ls Logo

GE LOGIQ P6

GE LOGIQ P6
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 37
8-2-6-3 Removal procedure (cont’d)
3.) Unscrew 2 screws from the rear side of the keyboard. Refer to the Figure 8-72.
4.) Lift the top cover of the keyboard and disconnect 4 cables from the keyboard. Refer to the Figure
8-73.
5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-6-4 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-72 Unscrews 2 screws
Figure 8-73 Lifting the top cover & Disconnecting 5 Cables
Table 8-7 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-6. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-7-1
Main Keyboard Assy Function Validation
Procedure
Section 4-3-4
System B/M-Mode Checks
Section 4-3-5
System CFM and PWD Checks
Section 10-5-2
Functional Checks (See Also Chapter 4)
(1) (2)

Table of Contents

Other manuals for GE LOGIQ P6

Related product manuals