DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
8 - 82 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-36 Probe Holder
8-2-36-1 Tools
• No need tool
8-2-36-2 Removal procedure
1.) Remove the probe holders from the Keyboard. Refer to the figure below.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-36-3 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-124 Probe Holders
Table 8-37 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-36. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-8-1
Cover Parts Function Validation
Section 10-5-5
Physical Inspection