DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
Section 8-4 - Software Installation 8 - 143
8-5-4 Restore procedure
1.) From Figure 8-140, press “Restore” button.
2.) Rest of the procedure is the same with “Backup” procedure.
3.) After reboot,
Press "Patient". Verify that Patient data are restored.
Press "Utility". Verify that Userdefined data are restored.
NOTICE
The system should be rebooted after finishing of Restore procedures.