Presets
LOGIQ V2/LOGIQ V1User Guide 3-27
Direction 5610736-100 Rev. 9
Restore procedure
1. Go to Utility --> System --> Backup/Restore.
2. In the Restore list,
• Select Patient Archive and Report Archive to restore
the patient archive.
• Select User Defined Configuration to restore all
system settings and user presets.
or
One or several system configuration items to restore
parts of the Detailed Restore of User Defined.
• Select Service to restore Service Settings.
3. In the Media field, select the appropriate Source device.
4. Select Restore.
The system performs the restore. As it proceeds, status
information is displayed on the Backup/Restore screen.
5. The LOGIQ V2/LOGIQ V1 restarts automatically when
Restore is done.
CAUTION
The restore procedure overwrites the existing database on the
local hard drive. Make sure to insert the correct media.
You cannot restore the data between systems with different
software versions.
CAUTION
To avoid the risk of overwriting the local patient and report
archives, DO NOT check Patient Archive when restoring
user-defined configurations.