System Start-Up
LOGIQ V2/LOGIQ V1 – User Guide 1-45
Direction 5610736-100 Rev. 9
System Start-Up
Connecting the System
Voltage level check
Check the rating label on the bottom of the system. Check the
voltage range indicated on the label.
CAUTION
Use caution to ensure that the power cable does not
disconnect during system use.
If the system is accidentally unplugged, data may be lost.
WARNING
Failure to provide an adequate earth circuit can cause electrical
shock, resulting in serious injury.
Connection of additional protective earth conductors or
potential equalization conductors is not necessary in most
cases and is only recommended for situations involving
multiple equipment in a high-risk patient environment to
provide assurance that all equipment is at the same potential
and operates within acceptable leakage current limits. An
example of a high-risk patient would be a special procedure
where the patient has an accessible conductive path to the
heart such as exposed cardiac pacing leads.
WARNING
To avoid risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
WARNING
POWER OUTAGE MAY OCCUR. The ultrasound unit requires
a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO
NOT have other equipment operating on the same circuit.

Table of Contents

Other manuals for GE LOGIQ V2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V2 and is the answer not in the manual?

GE LOGIQ V2 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals