Quality Assurance
LOGIQ V2/LOGIQ V1User Guide 3-69
Direction 5610736-100 Rev. 9
Ultrasound Quality Assurance Checklist
Table 3-10: Ultrasound Quality Assurance Checklist (Part 1)
Performed By Date
System Serial Number
Probe Type Probe Model Serial Number
Phantom Model Serial Number Room Temperature
Acoustic Output Gain Focal Zone
Gray Map TGC Depth
Monitor Setting
Peripheral Settings
Other Image Processing Control Settings
Table 3-11: (Part 2)
Test
Baseline Value Range
Tested Value
Image Hardcopy/Archived
Acceptable? Yes/No
Service Called (Date)
Date Resolved
Vertical Measurement Accuracy
Horizontal Measurement Accuracy
Axial Resolution
Lateral Resolution
Penetration
Functional Resolution
Contrast Resolution
Gray Scale Photography