EasyManua.ls Logo

GE LOGIQ V2

GE LOGIQ V2
381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Safety Precautions
LOGIQ V2/LOGIQ V1User Guide 4-29
Direction 5610736-100 Rev. 9
Best practices while scanning
NOTE: Refer to the Optimizing the Image sections for a complete
discussion of each control.
Acoustic Output Default Levels
In order to assure that an exam does not start at a high output
level, the LOGIQ V2/LOGIQ V1 initiates scanning at a reduced
default output level. This reduced level is preset programmable
and depends upon the exam category and probe selected. It
takes effect when the system is powered on or Patient is
selected.
To modify acoustic output, adjust the Power Output level.
HINTS
Raise the Acoustic Output only after attempting image
optimization with controls that have no effect on Acoustic
Output, such as Gain and TGC.
WARNING
Be sure to have read and understood control explanations for
each mode used before attempting to adjust the Acoustic
Output control or any control that can effect Acoustic Output.
Acoustic
Output
Hazard
Use the minimum necessary acoustic output to get the best
diagnostic image or measurement during an examination.
Begin the exam with the probe that provides an optimum focal
depth and penetration.

Table of Contents

Other manuals for GE LOGIQ V2

Related product manuals